Call Us Toll Free Email Us User Guides Sell Your Equipment
i-Cat Panoramic Dental Imaging System *Refurbished*
View Larger Image
i-Cat Panoramic Dental Imaging System *Refurbished*
SKU: 94
List Price: $89,995.00
   
Original Price: $79,995.00
Sale Price: $14,995.00
Your Savings: $75,000.00
Warranty: 90 Days Parts & Labor
   
Quantity:

2006 Model   /  i-Cat Classic Cone Beam Volumetric Tomography and Panoramic Dental Imaging System

CLICK HERE FOR INSTALLATION/SERVICE MANUAL

The system is a Cone Beam Volumetric Tomography and Panoramic X-ray device used for dental head and neck applications. The system consists of a Scanner and Computer Workstation which is suitable for an in-office environment. This scanning device is an open design that allows Patients to sit upright during a procedure. An electric powered seat is built into the device for proper Patient positioning.

The system captures data for 3D Skull Reconstruction for the following procedures:

• Implants
• TM Joints
• Reconstructed Panoramic
• Reconstructed Cephalometrics
• Airway / Sinus, etc.
• Nerve Canal 

Conventional Digital Panoramic Feature Cone Beam Volumetric Tomography is a medical imaging technique that uses X-rays to obtain cross-sectional images of the head or neck. Quality of the images depends on the level and amount of X-ray energy delivered to the tissue. Imaging displays both highdensity tissue, such as bone, and soft tissue. When interpreted by a trained Physician, these images provide useful diagnostic information. Intended Use of the Device The Imaging Sciences International (ISI) Scanner constructs a three dimensional model from images taken during a rotational X-ray sequence. The scanner is intended to be used whenever a dentist, oral surgeon, or other physician needs 3D information of high contrast objects. The system is designed for imaging of TM joint studies, mandible and maxilla for implant planning, sinuses, and other areas of the maxillofacial complex.

The system is comprised of the following major items:

• Scanner
• Computer CPU
• Operator Control Box with 50 ft. (15.2m) Cable
• Patient Emergency Stop Box with 10 ft. (3m) Cable
• Interlock Jumper Cable, 8 inch (20cm)
• Interlock Cable 50 ft. (15.2m)
• Warning Light Cable 50 ft. (15.2m)
• Chair Connection Cable, 10 ft. (3m)
• Power Cable, 15 ft.(4.6m)
• CAT 5 Ethernet Cable, 50 ft. (15.2m)

Technical Specifications

  • X-ray Source Tube Voltage: 120 kVp(eff)
  • Tube Current: 3-7 mA
  • Voltage Wave Shape: Constant Potential
  • Focal Spot: 0.0197 inches (0.5 mm)
  • Duty Cycle: 3%
  • Source to Sensor distance: 28.1 inches (71.4 cm)
  • Source to Patient distance*:19.5 inches (49.53 cm) (center of rotation) * The patient must be properly positioned in the Head Support Positioner Mechanism for each patient for all applications in order to have the focal spot to skin distance as large as possible.
  • Minimum Filtration (at 120 kVp(eff)) (mm of aluminum equivalent): 10 mm or greater
  • Maximum Rated Continuous Tube Operation: 130 kVp @ 0.5 mA
  • Maximum Rated Pulsed Tube Operation: 130 kVp @ 1mA
  • NOTE: Leakage technique factors are measured at the maximum specified energy.
  • Maximum Deviation: kV: +/- 5 kV
  • mA: + 10%
  • Timer: ± .01 seconds or 5%, whichever is greater
  • Maximum Excursion: 15 kV at 120 kV
  • X-ray Beam Size: Rectangular cone 9.37” wide x 1.97” to 7.56” high (23.8 cm width x 5 cm to 19.2 cm height)
  • PAN option: Rectangular cone 0.39” width x 6.30” height (1cm width x 16 cm height) (Automatically collimated not to exceed image detector readable area) 
  • Image Detector: Amorphous Silicon Flat Panel (readable area), 9.37” width x 7.56” height (23.8 cm width x 19.2 cm height)
  • Sensor Front Panel Attenuation Value: Less than 1mm of aluminum equivalent (information for reference only)
  • Gray Scale: 14 bit
  • Voxel Size: 0.4/0.3/0.25/0.2 mm
  • Image Acquisition: Single 360 degree rotation (maximum)
  • Scan Time: 26.9/8.9 seconds
  • Field of View: (standard) 6.50 inches diameter x 5.33 inches height (maximum) (16.5 cm diameter x 13.5 cm height (maximum)
  • Extended Field of View: (optional) 9.15 inches diameter x 6.67 inches height maximum) (23.2 cm diameter x 17 cm height (maximum))
  • Note: Maximum values can be collimated down.
  • Primary Reconstruction: Less than 2 minutes for 26.9 second scan @ 0.4 voxel
  • Secondary Reconstruction: Real Time
  • Stopping Distance and Angle: Hard stop is at -45o and 470o (reference is the gantry at the home position being 0o). Platform travel is 69 mm.

Laptop Minimum Requirements 

  • Intel Processor M 1.5Ghz
  • 1GB of RAM (2GB for peak performance)
  • Windows XP Professionalor Vista
  • 40GB Hard Drive
  • CDIDVD Drive
  • Integrated Gigabit Ethernet 10/100/100 Network Connection

Workstation Minimum Requirements

  • Processor 2Ghz
  • 1GB of RAM (2GB for peak performance)
  • Windows XP Professionalor Vista
  • Video Display capabilities of 1600 x 1200 resolution
  • 18 or 19 inch Monitor (recomended)
  • 40GB Hard Drive
  • CDIDVD Drive
  • Integrated Gigabit Ethernet 10/100/100 Network Connection

Weight

  • Total Weight: 510 lbs.
  • Tube Head Pod: 35.5 lbs.
  • Receptor Pod: 57 lbs.
  • X-Ray Power Supply: 9 lbs.